- Original article
- Open access
- Published:
Which stem in total hip arthroplasty for developmental hip dysplasia? A comparative study using a 3D CT-based software for pre-operative surgical planning
Journal of Orthopaedics and Traumatology volume 23, Article number: 33 (2022)
Abstract
Background
Stem choice in total hip arthroplasty (THA) for hip dysplasia is still controversial. The aims of the study were to evaluate (1) which stem design provided the highest percentage of adequate reconstructions in THA for dysplasia and (2) any correlation between the reconstructions provided by the stems and the native femoral morphology.
Materials and methods
150 CT scans including 200 adult dysplastic hips were randomly selected. Using the 3D CT-based software Hip-Op for surgical planning, the native hip anatomy was studied. Then, a single wedge tapered stem, an anatomical stem and a conical tapered stem were simulated in every hip. An adequate reconstruction of hip biomechanics was obtained when combined anteversion, offset restoration, coronal and sagittal tilt, canal filling and leg lengthening were inside the normal ranges.
Results
Conical stems achieved the highest percentage of adequate reconstructions (87%, p < 0.0001). The anatomical stem was the worst performer. Single wedge and anatomical stem acceptability was mainly influenced by the combined anteversion. Stem anteversion was correlated with the femoral anteversion (fair correlation), the calcar femorale (fair) and the mediolateral femoral diameter at isthmus (poor). When the femoral anteversion was ≥ 25°, combined anteversion was very acceptable for the conical stem (99.2%), whereas the rate of acceptable combined anteversion for the single wedge tapered stem was 71.4%, and that for the anatomical stem was 51.6% (p < 0.0001).
Conclusions
Stem choice in developmental hip dysplasia is mainly driven by appropriate combined anteversion, which is dependent on the coronal and axial femoral morphologies. As a rule of thumb, tapered stems are adequate when femoral anteversion is < 25°; conical stems should be adopted for higher anteversions.
Level of evidence
IV.
Introduction
The complex femoral morphology in developmental hip dysplasia (DDH) may pose several challenges to the reconstructive surgeons performing total hip arthroplasty (THA) [1,2,3,4,5,6]. In DDH, the main anatomic alterations are a high neck-shaft angle, a short offset, a narrow canal and, above all, a high femoral anteversion; these features, as well as the femoral anteversion, are not always correlated to the Crowe classification [1,2,3].
The wide variability and the complex geometry of dysplastic femurs make stem choice in THA critical to reconstructing hip biomechanics and confering implant stability [5]. To date, no guidelines and no thresholds based on the native morphology are currently available to facilitate a specific stem choice [4,5,6]. While some authors have empirically suggested conical tapered stems for femoral anteversions of 30–60° and for small femurs, the optimal stem design to use in DDH is still unknown [4,5,6].
Thus, we sought to determine (1) which stem design provided the highest percentage of adequate reconstructions in THA for DDH, in terms of combined anteversion, offset restoration, coronal and sagittal tilt, canal filling, and leg lengthening, and (2) any correlation between the reconstruction provided by each stem and the native femoral morphology. We hypothesized that in DDH, conical tapered stems could provide safe combined anteversion, reliable canal filling and adequate positioning in most cases, especially when the femoral anteversion is > 25°.
Materials and methods
The study was approved by the local ethical committee and registered on ClinicalTrials.gov (NCT04904640).
The hospital database was screened for pelvis CT scans that included a native dysplastic hip and were performed during 2000–2018 for pre-operative THA planning or painful THA diagnosis, with the aim being to achieve 200 eligible hips.
The inclusion criteria were:
-
Definition of DDH according to Wiberg (center edge angle < 20°) [7]
-
Pre-operative CT scan extended from the fourth lumbar vertebra to the tibial plateau.
Hips with other congenital or acquired pathologies, arthroplasties or inadequate CT scans were excluded.
200 hips in 150 CTs (150 Caucasian patients) were randomly selected. The first author simulated the implantation of three different stem designs per hip using a 3D CT-based pre-operative planning software (Hip-Op) after an appropriate assessment of the native femoral morphology.
3D CT-based pre-operative software for surgical planning: Hip-Op
The software reproduced a 3D CT-based planning environment [8, 9]. The user-friendly graphical user interface is based on a multimodal display visualization. The user can choose the components from a library of selected implants: the implants and the patient anatomy are rendered in each view. The planner may evaluate the implant type, size and position by interactively moving and rotating the components in the view area. The validity of Hip-Op software as a pre-operative planning aid has previously been assessed [9].
Implant features and planning technique
The planning was performed by a single arthroplasty surgeon (the first author) in every case. Three non-modular stems were selected based on their design and classification according to Khanuja et al. [10]: CLS (Zimmer Biomet, Warsaw, USA) single wedge tapered stem, type 1; Aptafix (Adler Ortho, Milan, Italy) anatomical stem, type 6; Wagner Cone (Zimmer Biomet) conical tapered stem, type 3B [11,12,13]. The CLS had 13 stem sizes with three CCD angles (125°, 135° and 145°) [11]. The Aptafix had a standard version and a offset solution (7.5 mm lateralization), both with a CCD angle of 135°: eight possible stem sizes are available [12]. The Wagner Cone had 12 stem sizes with two CCD angles, 125° and 135° [13]. The acetabular cup was the Continuum (Zimmer), which was adopted for every single stem simulation.
In all the dysplastic hips, the surgical planning aimed to reproduce the native center of rotation every possible time (a high hip center was only accepted in cases of very severe superolateral bone deficit). All the cups were implanted using a conservative circumferential reaming technique, admitting slight medialization in shallow acetabula. Cup inclination was set at 40–45°, with no more than one-third superior undercoverage. Cup anteversion was determined avoiding an anterior overhang (or reducing it to a minimum), aiming for 10–20° of anteversion. The stems were positioned with the aim of achieving an acceptable combined anteversion according to Dorr and Widmer, no offset reduction, adequate canal filling, the most neutral positioning in the coronal and sagittal planes, and leg length equalization, avoiding over-lengthening superior to 3 cm. For all the stems, the surgical planning was performed according to the instructions provided by the manufacturers and the experience of the planner.
Reliability of the simulations and comparison with post-operative measurements
Fifteen hips with a post-operative CT scan performed for contralateral hip planning and achieving a 2-year excellent clinical outcome with good signs of radiographic osseointegration were selected. In all cases, an anatomical stem was implanted. The first author, blinded to the post-operative component positioning, performed the simulation on the 15 hips. Another simulation was performed by the same author on the same hips after 4 weeks. All the tests were compared to the post-operative component positioning as provided by the CT scan. The outcomes and the reference values were the same as explained below.
Demographics of the cohort and assessment of the native hip morphology
The demographics of the patients were collected. On the CT scan, the DDH pathology was classified according to Crowe et al. and Hartofilakidis et al. [14]. The native femoral morphology was assessed in the coronal and axial planes [1,2,3, 15, 16] (Table 1).
Outcome measurements
In all the simulations, the following measurements were taken (cup measurements were unique to every case; stem measurements were performed for every stem in every case): cup anteversion, stem anteversion, acetabular offset, femoral offset, sagittal and coronal tilt, and canal filling at the mid-third of the stem [1, 11, 17, 18].
When planning, the following five outcomes had to be matched for the stem to be considered acceptable:
-
Combined anteversion (according to Dorr et al.) between 25° and 50° and combined anteversion (according to Widmer et al.) < 37° (both the targets had to be matched) [11];
-
Global offset (acetabular + femoral offset) loss not inferior to 12% of the native global offset [19];
-
Coronal and sagittal stem tilt < 5° [17];
-
Canal filling > 80% [18];
-
Leg lengthening < 3 cm [20].
Statistical analysis
The analyses were performed using SPSS 14.0 (SPSS Inc, Chicago, IL, USA). Quantitative data are reported as average values, standard deviations and minimum to maximum ranges. Qualitative data are expressed as frequencies and percentages and were tested using the chi-squared test. The reliability of the simulations was assessed using Fisher’s test (categorical variables). Correlations between parameters were assessed using Kendall, Spearman or Pearson coefficients, depending on the data type. The correlation strength was evaluated according to the current medical literature [21]. Threshold for significance: p = 0.05.
Results
Demographics and assessment of the native hip morphology
The cohort encompassed 35 males (23.3%) and 115 females (76.7%), with a mean age at the CT scan of 50.8 ± 11 years (range: 19–81). The mean height was 159.8 ± 18.3 cm (range: 128–185), the mean weight was 66.1 ± 15.3 kg (range: 40–112): the average BMI was 25.8 ± 4.9 kg/m2 (range: 17.2–39.6). According to the Crowe classification, 92 hips (42%) were graded as I, 62 (32%) as II, 37 (18.5%) as III and 9 (4.5%) as IV. According to the Hartofilakidis classification, 97 hips (48.5%) were classified as A, 59 (29.5%) as B1, 35 (17.5%) as B2, 2 (1%) as C1 and 7 (3.5%) as C2. Both classifications were statistically correlated (Kendall τ = 0.956, p < 0.0001). The features of the native hip morphology and the correlations with the Crowe classification are provided in Table 1. The center of rotation height progressively increased with the Crowe classification (moderate correlation), as well as the femoral anteversion (fair), the leg length discrepancy (fair) and the angle between the calcar femorale and the tangent to the posterior condyles (poor). The femoral offset (fair correlation), the mediolateral femoral diameter 2 cm above the lesser trochanter (fair), the canal flare index (fair) and the acetabular anteversion (poor) progressively decreased with increasing DDH degree (Crowe classification). The same correlations with the same coefficients were evident when the Hartofilakidis classification was adopted.
Reliability of the simulations
There was perfect agreement between the simulations and the post-operative CT measurements for combined anteversion, leg lengthening, coronal tilt, sagittal tilt and canal filling. There was no significant difference in offset restoration (Fisher's test, p = 0.141). Perfect intra-observer reliability was recorded for all six measured parameters.
Percentages of the implants that provide adequate reconstruction for each stem design
The stem design providing the highest percentage of adequate hip reconstruction (i.e., all six parameters were matched) was the conical implant (87% of the cases, chi-square test, p < 0.0001) (Fig. 1). Out of the 13% of conical implants that did not achieve adequate reconstruction, more than half of those cases did not show good offset restoration (7% of all the implants) (chi-square test, p < 0.0001). Single wedge and anatomical stems could not be simulated due to abnormal femoral anatomy in 4.5% and 0.5% of the cases, respectively. Conical implants matched the combined anteversion target in almost all cases; the anatomical stem was the worst performer (Table 2).
Correlations between the reconstructions provided by the stem designs and the native morphology
The reconstructions provided by the stem designs were significantly correlated with the Crowe classification. For Crowe I and II hips, the single wedge implant provided an acceptable reconstruction in 70.8% of the cases, the anatomical stem in 44.8% and the conical device in 91.6% (Pearson’s chi-square test, p < 0.0001) (Fig. 2). For Crowe III and IV hips, the same designs achieved percentages of acceptability of 63%, 41.3% and 71.7%, respectively (Pearson’s chi-square test, p < 0.0094) (Fig. 3). Similar percentages were achieved with the Hartofilakidis classification, apart from C hips (the single wedge and conical stem performed equally well, p 0.837).
The acceptability of the single wedge stems was mainly correlated with combined anteversion (Pearson’s r = 0.827, p < 0.0001: very strong correlation), leg lengthening < 3 cm (Pearson’s r = 0.377, p < 0.0001: fair correlation) and offset reconstruction (Pearson’s r = 0.308, p < 0.0001: fair correlation) (Table 3). The acceptability of the anatomical stems was mainly correlated to combined anteversion (Pearson’s r = 0.716, p < 0.0001: moderate correlation), offset reconstruction (Pearson’s r = 0.415, p < 0.0001: fair correlation), canal filling (Pearson’s r = 0.320, p < 0.0001: fair correlation) and leg lengthening < 3 cm (Pearson’s r = 0.270, p 0.0001: fair correlation). For the conical stems, the acceptability was correlated to offset reconstruction (Pearson’s r = 0.710, p < 0.0001: moderate correlation), leg lengthening < 3 cm (Pearson’s r = 0.455, p < 0.0001: fair correlation), combined anteversion (Pearson’s r = 0.319, p < 0.0001: fair correlation) and canal filling (Pearson’s r = 0.26, p < 0.0002: fair correlation).
In general, stem anteversion was correlated with femoral anteversion (Pearson’s r = 0.380, p < 0.0001: fair correlation), the angle between the calcar femorale and the posterior femoral condyles (Pearson’s r = 0.284, p < 0.0001: fair correlation), and the mediolateral femoral diameter at isthmus (Pearson’s r = − 0.133, p 0.0012: poor correlation). When the femoral anteversion was ≥ 25°, combined anteversion acceptability was 99.2% for the conical stem, 71.4% for the single wedge stem and 51.6% for the anatomical stem (Pearson’s chi-square test, p < 0.0001) (Table 4). The same trend with femoral anteversion was observed for acceptable reconstructions based on all six criteria. For offset reconstruction, all the stem designs provided adequate percentages (single wedge 96.9%; anatomical 84.6%; conical 90.8%) of offset restoration when the native offset was normal (37 ± 4 mm), with no significant differences between the stems (Pearson’s chi-square test, p = 0.056) [19]. When the native offset was beyond the normal limits, the single wedge stem achieved the best performance (95.6%, Pearson’s chi-square test, p < 0.0001), followed by the conical (94.1%) and anatomical (80.7%) stems.
Discussion
Stem choice in DDH is still a matter of debate [6]. There are no available guidelines, only generic recommendations for conical implants in cases with small femurs, high anteversions and abnormal anatomies [4,5,6]. The study showed that conical stems provided the best percentage of acceptable implants in DDH. Stem choice was mainly dictated by combined anteversion; in particular, stem anteversion was correlated with the native femoral anteversion, the calcar femorale and the mediolateral diameter of the femur. Conical stems adequately reconstructed the hip biomechanics, even in the case of high femoral anteversion, ≥ 25°.
The simulation of three different stem designs in the same 200 randomly selected hips provided a valuable tool to define appropriate indications and thresholds for stem choice in DDH. However, this study has some methodological limits: the single-surgeon planning technique, the variability of femoral anteversion measurements and the small cohort of high-degree dysplasia. On the other hand, as shown by the description of the native dysplastic morphology, a wide range of femoral variations and axial anatomies were included in the study. Some exceptional cases were probably excluded or poorly represented; however, THA simulations of outliers are beyond the scope of the study. A single stem simulation per stem group was provided, according to the available software library, the collaboration of manufacturers and the simulator’s proficiency with the specific stem. The simulated reconstructions provided by each stem should be considered indicative of any other stem belonging to the same group: minor differences due to the variability of the proximal biomechanics of other similar stems do not invalidate the general conclusions. Modular stems were intentionally excluded: these devices are valid alternatives in complex anatomies but at the cost of possible adjunctive failures, especially in young and active males [20].
From a practical perspective, the study showed that single wedge biomechanical reconstructions in DDH were mainly influenced by the combined anteversion, and thus benefited from lower native femoral anteversion and a lower calcar femorale angle. Similarly, anatomical stem acceptability was impacted by the combined anteversion and consequently was favored by lower native femoral anteversion. Single wedge stems matched the correct combined anteversion in more cases than anatomical stems, confirming that tapered stems may allow a larger anteversion adjustment than metaphyseal fitting stems [5]. However, single wedge stems could not address as many femoral deformities as conical implants, which matched the correct combined anteversion in almost all cases (> 97%) regardless of the native femoral anteversion, thanks to the free adjustment of the version [6]. Conversely, most of the conical stems that did not fulfill the criteria of good reconstruction in DDH failed due to inadequate offset restoration (7%), highlighting a well-known drawback of these implants [6]. Thus, considering the good percentage of hip reconstruction, with only a small percentage of cases showing non-optimal offset restoration, it could be concluded that conical stems should be recommended for most (all) dysplastic hips. However, considering that conical stems have a diaphyseal fitting design that violates the distal bone stock, and offset reconstruction may be impaired in low-degree DDH, the routine adoption of conical stems in DDH cases does not seem warranted [6] (Table 1).
Thus, we could assume that the pre-operative planning of THA in DDH should play a pivotal role in discriminating between stems. However, the sole coronal images with Crowe and Hartofilakidis classifications are not informative enough to allow a mindful choice in every case, as bidimensional classifications have only a fair correlation with axial femoral parameters, as demonstrated by the standard deviations and the medians of the four Crowe cohorts (Table 1). As adequate reconstruction of hip biomechanics is mainly based on appropriate combined anteversion and stem anteversion, which depends on the mediolateral diameter at the isthmus, the calcar femorale and the femoral anteversion, axial imaging seems to be of help when choosing the most appropriate stem.
Thus, it is difficult to provide a simple femoral classification for stem choice in DDH, as the native femoral morphology develops along the coronal and the axial planes simultaneously. A pre-operative CT scan with dedicated protocols may allow comprehensive planning (with no more concerns about radiation exposure) [22]. However, as a rule of thumb, tapered stems can provide adequate reconstruction in cases with femoral anteversion < 25° (thus, in most low-degree DDH), and have the advantage of respecting the diaphyseal bone stock and producing the best offset reconstruction. Conical stems should be implanted in cases with femoral anteversion ≥ 25°, but also in cases with small femurs and abnormal anatomies regardless of the magnitude of femoral anteversion.
Data availability
The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.
References
Sugano N, Noble PC, Kamaric E, Salama JK, Ochi T, Tullos HS (1998) The morphology of the femur in developmental dysplasia of the hip. J Bone Joint Surg Br 80(4):711–719
Argenson JN, Ryembault E, Flecher X, Brassart N, Parratte S, Aubaniac JM (2005) Three-dimensional anatomy of the hip in osteoarthritis after developmental dysplasia. J Bone Joint Surg Br 87(9):1192–1196
Argenson JN, Flecher X, Parratte S, Aubaniac JM (2007) Anatomy of the dysplastic hip and consequences for total hip arthroplasty. Clin Orthop Relat Res 465:40–45
Sugano N, Takao M, Sakai T, Nishii T, Miki H (2012) Does CT-based navigation improve the long-term survival in ceramic-on-ceramic THA? Clin Orthop Relat Res 470(11):3054–3059
Taniguchi N, Jinno T, Koga D, Hagino T, Okawa A, Haro H (2017) Cementless hip stem anteversion in the dysplastic hip: a comparison of tapered wedge vs metaphyseal filling. J Arthroplasty 32(5):1547–1552
Zhang Q, Goodman SB, Maloney WJ, Huddleston JI 3rd (2016) Can a conical implant successfully address complex anatomy in primary THA? radiographs and hip scores at early followup. Clin Orthop Relat Res 474(2):459–464
Cooperman D (2013) What is the evidence to support acetabular dysplasia as a cause of osteoarthritis? J Pediatr Orthop 33(Suppl 1):S2-7
Rizzoli Orthopedic Institute (2020) Hip-Op software. Rizzoli Orthopedic Institute, Bologna. http://www.hipop-plan.com/. Accessed 28 Oct 2020
Viceconti M, Lattanzi R, Antonietti B, Paderni S, Olmi R, Sudanese A, Toni A (2003) CT-based surgical planning software improves the accuracy of total hip replacement preoperative planning. Med Eng Phys 25(5):371–377
Khanuja HS, Vakil JJ, Goddard MS, Mont MA (2011) Cementless femoral fixation in total hip arthroplasty. J Bone Joint Surg Am 93(5):500–509
Zimmer Biomet (2020) CLS stem. Zimmer Biomet, Warsaw. https://www.zimmer.co.uk/content/dam/zimmer-web/documents/en-GB/pdf/medical-professionals/hip/06-01073-012X-000-01.pdf. Accessed 28 Oct 2020
Adler Ortho (2020) Aptafix stem. Adler Ortho, Milan. http://www.adlerortho.com/wp-content/uploads/2020/02/TO-APTA-FIX-SURG-TEC-ITA-2019-1.pdf. Accessed 28 Oct 2020
Zimmer Biomet (2020) Wagner Cone stem. Zimmer Biomet, Warsaw. https://www.zimmer.pt/content/dam/zimmer-web/documents/en-US/pdf/surgical-techniques/hip/wagner-cone-prosthesis-hip-stem-surgical-technique.pdf. Accessed 28 Oct 2020
Decking R, Brunner A, Decking J, Puhl W, Günther KP (2006) Reliability of the Crowe und Hartofilakidis classifications used in the assessment of the adult dysplastic hip. Skeletal Radiol 35(5):282–287
Castagnini F, Valente G, Crimi G, Taddei F, Bordini B, Stea S, Toni A (2019) Component positioning and ceramic damage in cementless ceramic-on-ceramic total hip arthroplasty. J Orthop Sci 24(4):643–651
Tetsunaga T, Fujiwara K, Endo H, Tetsunaga T, Shiota N, Sato T, Ozaki T (2017) Calcar femorale in patients with osteoarthritis of the hip secondary to developmental dysplasia. Clin Orthop Surg 9(4):413–419
García-Rey E (2021) The fate of sagittal alignment in tapered uncemented femoral components in total hip arthroplasty: 889 hips followed for a minimum of 10 years. Hip Int 31(4):472–481
Kobayashi K, Kidera K, Itose M, Motokawa T, Chiba K, Osaki M (2020) Higher incidence of aseptic loosening caused by a lower canal filling ratio with a modified modular stem in total hip arthroplasty. J Orthop Surg Res 15(1):568
Asayama I, Chamnongkich S, Simpson KJ, Kinsey TL, Mahoney OM (2005) Reconstructed hip joint position and abductor muscle strength after total hip arthroplasty. J Arthroplasty 20:414–420
Takao M, Ohzono K, Nishii T, Miki H, Nakamura N, Sugano N (2011) Cementless modular total hip arthroplasty with subtrochanteric shortening osteotomy for hip with developmental dysplasia. J Bone Joint Surg Am 93:548–555
Akoglu H (2018) User’s guide to correlation coefficients. Turk J Emerg Med 18(3):91–93
Salem HS, Marchand KB, Ehiorobo JO, Tarazi JM, Matzko CN, Sodhi N, Hepinstall MS, Mont MA (2020) Benefits of CT scanning for the management of hip arthritis and arthroplasty. Surg Technol Int 36:364–370
Acknowledgements
Not applicable.
Location statement
The study was performed at Istituto Ortopedico Rizzoli, Bologna, Italy.
Funding
No funding was received.
Author information
Authors and Affiliations
Contributions
FC: conceptualization, methodology, investigation, writing. SL and FP: resources, software, investigation. BB and MC: formal analysis and data curation. FT: conceptualization, project administration, validation visualization. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
Institutional review board approved the study (IRB 29/2021/Oss/IOR, 1st February 2021).
Consent for publication
Not applicable.
Competing interests
The authors declare that they have no competing interests.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.