Arthrodesis of proximal inter-phalangeal joint for hammertoe: intramedullary device options

Abstract

Background

Proximal inter-phalangeal (PIP) joint arthrodesis today represents the standard treatment for structured hammertoes; however, recently, a lot of new intramedullary devices for the fixation of this arthrodesis have been introduced. The purpose of this work is to look at the currently available devices and to perform a review of the present literature.

Materials and methods

A literature search of PubMed/Medline and Google Scholar databases, considering works published up until September 2014 and using the keywords: hammertoe, arthrodesis, PIP joint, fusion, intramedullary devices, and K-wire, was performed. The published papers were included in the present study only if they met the following inclusion criteria: English articles, arthrodesis of PIP joints for hammertoes with new generation intramedullary devices, series with n > 10. Studies using absorbable pins or screws that are considered as another kind of fixation that involved more than one articulation, as well as comments, letters to the editor, or newsletters were excluded.

Results

Nine publications were included. Of the patients’ reports, 93–100 % were good or excellent concerning satisfaction. Radiological arthrodesis was achieved in 60.5–100 % of cases. Three of the publications compared the new devices with the K-wire. Of these three articles, two employed the traditional technique and one the buried technique. The AOFAS score, evaluated in three publications, showed a delta of 19, 45 and 58 points. Major complications, which required a secondary surgical revision, were between 0 and 8.6 %. The complications of the K-wire and the new devices were similar; also the reoperation rate was close to equal (maximal difference 2 %). On the other hand, these kinds of devices definitely have a higher price, compared to the K-wire.

Conclusion

The use of these new devices provides good results; however, their high price is currently a problem. For this reason, cost-benefit studies seem to be necessary to justify their use as standard treatment.

Level of evidence

Level III systematic review.

Introduction

Nowadays, the treatment of the hammertoe is still disputable; indeed, a lot of procedures, both on the soft tissues and the bone structures are purposed and considered efficient. In the rigid and structured deformities not suited for manual correction, arthrodesis of the proximal inter-phalangeal (PIP) joint represents the standard treatment [1]. This procedure is performed by removing the articular surfaces of the proximal and intermediate phalanges. Many systems such as cannulated screws or absorbable pins have been designed for the fixation of the arthrodesis, yet still the K-wire is the traditional method, and most utilized [26]. However, recently, new intramedullary devices have been used persistently, trying to solve problems such as infections [3, 7], traumatic breaks [8, 9] and malalignments [10] tied to the K-wire.

As of today (September 2014), after accurate research, 16 different devices are available on the United States (US) and European (EU) markets. These were divided into four categories according to technical features and material composition (Table 1).

Table 1 Intramedullary devices available on US and EU markets (up to September 2014)
  • Shape memory devices: these are composed of a memory metal (Memometal NiTinol), which is activated by body temperature, modifying its shape once implanted. Specifically, these become shorter and enlarge themselves to bestow more stability to the system.

  • Bone allograft devices: since these devices are grafts, they have bone inductive and conductive properties, which improve their integration significantly.

  • One-piece solid or cannulated devices: thanks to the form of their extremities, these can be anchored to the cortical of the proximal and middle phalanges. The cannulated type also permits the use of the K-wire as a guide. With these devices the proximal part is threaded and screwed onto the proximal phalange, while the distal part is anchored to the middle phalange. They are available in steel, titanium or polyetheretherketone (PEEK).

  • Two-piece devices: a female and a male part make up these devices. Once positioned, one on the proximal phalanges, and the other one on the middle, these are fixed together.

The purpose of this work is to look at the currently available devices and to review, from the literature, the results of these for PIP fusion.

Materials and methods

A literature search of PubMed/Medline and Google Scholar databases, considering works published up until September 2014, and using the keywords: hammertoe, arthrodesis, PIP joint, fusion, intramedullary devices, and K-wire was performed. The published papers were included in the present study only if they met the following inclusion criteria: English articles, arthrodesis of PIP joints for hammertoes with new generation intramedullary devices, series with n > 10. Studies using absorbable pins or screws that are considered as another kind of fixation that involved more than one articulation, as well as comment, letter to editor or newsletters were excluded.

The search strategy identified over 455 articles. A total of 43 publications describing specifically the arthrodesis of the PIP joint for hammertoe could be identified.

Thirty-four articles were excluded due to exclusion criteria, these being: studies using absorbable pins, screws or other kinds of fixation (n = 23), fewer than ten patients (n = 9), non-English language (n = 1), comment (n = 1) (Fig. 1).

Fig. 1
figure 1

Flowchart of eligible study evaluation

Nine articles, finally, met the inclusion criteria and were compatible with our review (Table 2) [5, 1017]. In these papers, patient satisfaction, achievement of arthrodesis, AOFAS score and the rate of complications were taken into consideration.

Table 2 Results of the reviewed articles

Results

The results from the nine articles included in the work are reported in Table 2.

The satisfaction of the patient, taken into consideration in four publications, reports a good/excellent result in 93–100 % of the cases [1113, 15].

In contrast, radiological arthrodesis is achieved in 60.5–100 % of the cases [4, 11]. This value result is heterogeneous and is barely correlated to the review cases, demonstrating the frequent establishment of a fibrous union.

Three publications compare the new devices to the K-wire: two of these use the traditional technique and one the buried technique. The Angirasa et al. and Roukis et al. publications report more satisfaction for the devices, yet none of these works cite any cases of revision [10, 11]. The Scholl et al. [17] group, instead, reports no significant difference of revisions utilizing K-wire with the buried technique (8.6 % against 10.7 % p = 0.754).

The AOFAS score (evaluated in three publications) shows a delta of 19, 45 and 58 points [12, 13, 15].

Minor complications, often asymptomatic and radiologically identified have been: malunion (2.4–7 %) [5, 10], displaced fixation (1.5–13 %) [10], mallet toe (2–23 %) [10, 13], non union (1.5 %) [16], hardware failures (3–5 %) [12, 16] and ruptures (5 %) [12].

The major complications, which required a surgical revision, vacillate between 0 and 8.6 %. These complications were mainly due to malunion, breaks or recurrence.

In conclusion, only two works took the price of the devices into consideration; Coillard et al. reported a 20 times higher price of these devices compared to the K-wire [13]. Ellington et al., instead, reported a price of $225 per device (StayFuse™, Nexa Orthopaedics, San Diego, CA) [14].

Discussion

Although hammertoe is a very frequent disease, the treatment is still heavily disputed. In the structured deformities not suited for manual correction, PIP fusion is considered, today, the standard treatment [4]. The K-wire technique is the most utilized method for performing the fusion, as it is fast, cheap and simple to implant [3]. On the other hand, this kind of fixation method also has weak points: the exterior communication that predisposes for infections and traumatisms, the violation of the distal inter-phalangeal (DIP) joint, the lack of compression and rotational control and, finally, discomfort at removal [3, 710].

Because of this, the intramedullary devices aim to solve the weak points of the K-wire technique. Indeed, the results reported above seem to be slightly better than those of the K-wire, especially regarding patient satisfaction and malalignment of the arthrodesis.

Considering everything, the type of complications reported for the new devices and the K-wire treatment have been similar, save the superficial infections. Taking into consideration the major complications, in other words the cases which needed a reoperation in the articles that directly compare the new devices to the K-wire, no differences were found [10, 11], or in any case no statistically significant differences [17].

On the other hand, as reported by Ellington and Coillard, the devices definitely have higher prices compared to the K-wire, which represents a limit to their utilization, especially in the case of multiple toe corrections [13, 14].

Currently, no evidence exists in the literature which justifies the use of these new devices, especially considering their high price. For this reason cost-benefit studies are necessary to understand whether lower reoperation rates can justify the use of these devices as the new standard treatment in the future for hammertoes.

Regarding reoperation, this can also result in difficulties, especially in the phase of the removal of the device, and cause an excessive reduction of the toe length. For this reason new materials such as PEEK aim to make the revision easier.

Of the 16 devices currently available on the US and EU markets, as reported in Table 2, only four are also described in the literature (according to the criteria previously mentioned) and did not show significant differences in their results. For the remaining devices, future studies are still necessary.

In conclusion, the new intramedullary devices represent an interesting topic because of their continuous evolution and the constant birth of new devices on the market with new characteristics and material compositions.

The use of these devices seem to provide good results; however, the dilemma tied to their high price is not negligible. For this reason, cost-benefit studies that are still lacking in the literature seem necessary to justify the supremacy and the use of the new devices in the future as standard treatment for hammertoes.

References

  1. Zelen CM, Young NJ (2013) Digital arthrodesis. Clin Podiatr Med Surg 30(3):271–282. doi:10.1016/j.cpm.2013.04.006

    Article  PubMed  Google Scholar 

  2. Caterini R, Farsetti P, Tarantino U, Potenza V, Ippolito E (2004) Arthrodesis of the toe joints with an intramedullary cannulated screw for correction of hammertoe deformity. Foot Ankle Int 25(4):256–261. doi:10.1177/107110070402500411

    CAS  PubMed  Google Scholar 

  3. Coughlin MJ, Dorris J, Polk E (2000) Operative repair of the fixed hammertoe deformity. Foot Ankle Int 21(2):94–104. doi:10.1177/107110070002100202

    CAS  PubMed  Google Scholar 

  4. Ellington JK (2011) Hammertoes and clawtoes: proximal interphalangeal joint correction. Foot Ankle Clin 16(4):547–558. doi:10.1016/j.fcl.2011.08.010

    Article  PubMed  Google Scholar 

  5. Scott RT, Hyer CF, Berlet GC (2013) The PROTOE intramedullary hammertoe device: an alternative to Kirschner wires. Foot Ankle Spec. 6(3):214–216. doi:10.1177/1938640013487891

    Article  PubMed  Google Scholar 

  6. Smith BW, Coughlin MJ (2009) Disorders of the lesser toes. Sports Med Arthrosc 17(3):167–174. doi:10.1097/JSA.0b013e3181a5cd26

    Article  PubMed  Google Scholar 

  7. Klammer G, Baumann G, Moor BK, Farshad M, Espinosa N (2012) Early complications and recurrence rates after Kirschner wire transfixion in lesser toe surgery: a prospective randomized study. Foot Ankle Int 33(2):105–112. doi:10.3113/FAI.2012.0105

    Article  PubMed  Google Scholar 

  8. Konkel KF, Sover ER, Menger AG, Halberg JM (2011) Hammer toe correction using an absorbable pin. Foot Ankle Int 32(10):973–978. doi:10.3113/FAI.2011.0973

    Article  PubMed  Google Scholar 

  9. Zingas C, Katcherian DA, Wu KK (1995) Kirschner wire breakage after surgery of the lesser toes. Foot Ankle Int 16(8):504–509. doi:10.1177/107110079501600809

    Article  CAS  PubMed  Google Scholar 

  10. Roukis TS (2009) A 1-piece shape-metal Nitinol intramedullary internal fixation device for arthrodesis of the proximal interphalangeal joint in neuropathic patients with diabetes. Foot Ankle Spec 2(3):130–134. doi:10.1177/1938640009336199

    Article  PubMed  Google Scholar 

  11. Angirasa AK, Barrett MJ, Silvester D (2012) SmartToe® implant compared with Kirschner wire fixation for hammer digit corrective surgery: a review of 28 patients. J Foot Ankle Surg 51(6):711–713. doi:10.1053/j.jfas.2012.06.013

    Article  PubMed  Google Scholar 

  12. Catena F, Doty JF, Jastifer J, Coughlin MJ, Stevens F (2014) Prospective study of hammertoe correction with an intramedullary implant. Foot Ankle Int 35(4):319–325. doi:10.1177/1071100713519780

    Article  PubMed  Google Scholar 

  13. Coillard JY, Petri GJ, Van Damme G et al (2014) Stabilization of proximal interphalangeal joint in lesser toe deformities with an angulated intramedullary implant. Foot Ankle Int 35(4):401–407. doi:10.1177/1071100713519601

    Article  PubMed  Google Scholar 

  14. Ellington JK, Anderson RB, Davis WH, Cohen BE, Jones CP (2010) Radiographic analysis of proximal interphalangeal joint arthrodesis with an intramedullary fusion device for lesser toe deformities. Foot Ankle Int 31(5):372–376. doi:10.3113/FAI.2010.0372

    Article  PubMed  Google Scholar 

  15. Fazal MA, James L, Williams RL (2013) StayFuse for proximal interphalangeal joint fusion. Foot Ankle Int 34(9):1274–1278. doi:10.1177/1071100713485545

    Article  CAS  PubMed  Google Scholar 

  16. Sandhu JS, DeCarbo WT, Hofbauer MH (2013) Digital arthrodesis with a one-piece memory Nitinol intramedullary fixation device: a retrospective review. Foot Ankle Spec 6(5):364–366. doi:10.1177/1938640013496458

    Article  PubMed  Google Scholar 

  17. Scholl A, McCarty J, Scholl D, Mar A (2013) Smart toe® implant versus buried Kirschner wire for proximal interphalangeal joint arthrodesis: a comparative study. J Foot Ankle Surg 52(5):580–583. doi:10.1053/j.jfas.2013.02.007

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Matteo Guelfi.

Ethics declarations

Conflict of interest

The authors declare that they have no conflict of interest. The authors did not receive any outside funding or grants in support of their research for or preparation of this work. No commercial entity paid or directed, or agreed to pay or direct, any benefits to any research fund, foundation, division, center, clinical practice, or other charitable or non-profit organization with which the authors, or a member of their immediate families, are affiliated or associated.

Ethical standards

This article does not contain any studies with human participants performed by any of the authors.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Guelfi, M., Pantalone, A., Daniel, J.C. et al. Arthrodesis of proximal inter-phalangeal joint for hammertoe: intramedullary device options. J Orthopaed Traumatol 16, 269–273 (2015). https://doi.org/10.1007/s10195-015-0360-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10195-015-0360-0

Keywords