FiberWire tension band for patellar fractures

Abstract

Background

Symptomatic hardware represents the most frequent complication reported following surgical treatment of patellar fracture. For this reason, some authors suggested using nonabsorbable sutures to fix the fracture with various techniques. The aim of this study was to evaluate clinical and radiological results of patients treated following a modified Pyriford technique using a FiberWire suture (Arthrex, Naples, FL, USA).

Materials and methods

We retrospectively evaluated a case series of  seventeen patients with displaced patellar fractures treated by open reduction and internal fixation with a modified tension band using FiberWire sutures. Clinical and radiological outcome were evaluated. Union time, complications, and reoperation rate were observed and recorded.

Results

All fractures healed (time to union 9.2 ± 2 weeks), and no fixation failure was observed. Slight losses of reduction (<4 mm) were noted in two patients at 4 weeks postoperatively. The average Lysholm and Bostman scores at the final follow-up were 91 ± 5.7 (range 83–100) and 28.3 ± 1.6 (range 26–30), respectively.

Conclusion

Modified tension band using FiberWire sutures showed satisfactory clinical results, with a low incidence of complications and reoperations. FiberWire tension bands could be used in place of metal-wire tension bands to treat patellar fracture, reducing the rate of symptomatic hardware.

Level of evidence

4

Introduction

Patellar fractures are relatively uncommon, representing approximately 1% of all skeletal injuries [2]. Nonsurgical management is recommended for fractures with intact extensor mechanism, minimal intra-articular stepoff, and minimal fracture displacement (1–4 mm) [3]. An incompetent extensor mechanism associated with a displaced or comminuted fracture with a torn extensor retinaculum is indicated for surgery. Operative treatment must provide stable patellar fracture fixation to allow early mobilization and prevent fracture displacement. Furthermore, articular congruity is essential to reduce the increased risk of posttraumatic osteoarthritis as a result of the high-contact forces in the patellofemoral joint [5].

Several different techniques for internal fixation have been proposed and employed over the years, with different rates of success. Historically, the most commonly used technique for managing patellar fracture fixation is represented by the modified tension-band wiring technique [5, 13, 19, 23]. It involves longitudinal Kirschner wires (K-wires) and 18-gauge stainless steel wire in a figure-of-eight pattern looped over the anterior side of the patella. This technique neutralizes tension forces anteriorly produced by the extensor mechanism at knee flexion and converts them into stabilizing compressive forces at the articular surface [23]. This construct represents the most widely used method of fixation for transverse and comminuted patellar fracture. Over the years, this technique was further modified by different authors using either K-wires or cannulated screws with different stainless steel wire configurations. Even though substantially good clinical results have been reported, all techniques using metallic wires can result in symptomatic hardware, which represents the most frequent complication, with reported rates varying from 0 % to 60 % [12, 14, 20]. In fact, implants can irritate overlying soft tissue and cause pain, requiring a second surgery to remove the implant. In addition, a high incidence of infection has been reported [9, 12]. For this reason some authors have advocated the use of nonabsorbable sutures, such as braided polyester, to fix the fracture when using various techniques [6, 7, 9, 12, 21]. Several advantages of using a suture over a wire have been shown in the literature, such as lower rate of revision surgery and higher patient tolerance of surgical material that does not irritate soft tissues as much as wire. Another advantage is represented by easier handling of suture materials for the surgeon, which could determine shorter operating and tourniquet time [9, 11]. Few authors reported data concerning the use of tension-band fixations employing nonabsorbable sutures. Although no. 5 Ethibond (Ethicon, Somerville, NJ, USA) and no. 5 Ti-Cron (Davis and Geck, Gosport, Hampshire, UK) braided polyester sutures have been clinically and biomechanically studied for patellar fracture fixation, to our knowledge there are no data on the use of FiberWire sutures (Arthrex, Naples, FL, USA).

The aim of this study was to evaluate clinical outcome and rate of postoperative complications of patients who underwent an open reduction and internal fixation (ORIF) using a tension-band wiring technique with a FiberWire suture. The authors hypothesized that a FiberWire tension band could be used for patellar fracture fixation, decreasing the rate of symptomatic hardware without reducing the rate of fracture healing.

Materials and methods

We performed a retrospective study of consecutive patients hospitalized for patellar fractures in our Institution (Orthopaedic and Traumatology Unit, AOUP “P.Giaccone”, Palermo, Italy) from 2008 to 2013. Inclusion criteria for the study were the use of FiberWire suture for fracture fixation, age ≥18 years, and minimum follow-up of at least 10 months. Exclusion criteria were open fractures, stainless steel wire implantation, previous ipsilateral knee surgery, and patients with polytrauma or head injuries that definitely influence rehabilitation.

Fifty-one patients with patellar fractures were admitted to the hospital during the mentioned period. Among them, 37 required ORIF. Seventeen patients were excluded: specifically, eight were treated with a suture other than the FiberWire; in six patients, a hybrid technique was used (metallic K-wires and nonabsorbable suture); two patients presented and open fracture; one patient presented with polytrauma that required long hospitalization, compromising the rehabilitation program. Twenty patients met inclusion criteria. All patients were contacted by telephone and/or email and were asked to return to the hospital for a final physical examination. In order to maximize follow-up, Internet-based searches was used to contact patients who did not respond to the first mailing. Three patients were lost to follow-up: one patient refused the final clinical assessment because she was living abroad; two patients could not be located. For all patients, medical records and X-ray films were reviewed and fracture types classified according to the Orthopaedic Trauma Association fracture classification (AO/OTA) system.

In all patients, surgery was scheduled 2–3 days after tje trauma and was performed by the senior author. Patients were operated on in a supine position, and a tourniquet was placed high up on the thigh. In all cases, a longitudinal skin incision on the knee was performed. An extemporary reduction was performed with a reduction clamp and checked under fluoroscopy. In this phase, care was taken to obtain the best possible reduction and congruence of the articular patellar surface. At this point, a modified Pyriford technique was used for fracture fixation [8]. A no. 5 FiberWire nonabsorbable suture was used, and a peripatellar circumferential cerclage was performed in a purse-string fashion close to the bone. This allowed initial fracture stabilization. A second no. 5 FiberWire suture was then placed in a figure-of-eight fashion through the quadriceps and patellar tendon to obtain a modified anterior tension band. Each of the two FiberWire bands was manually tensioned and knotted on the superior board of the patella (Fig. 1). Additionally, retinacular defect was repaired with an absorbable suture.

Fig. 1
figure 1

FiberWire tension band. Peripatellar circumferential cerclage (a); modified anterior tension band (b); final construct (c)

All patients were immobilized postoperatively with a long-hinged knee brace locked at 0° for 3 weeks. Weight bearing started as tolerated with crutches, while full weight bearing was allowed from the fourth week postoperatively. Passive range of motion (ROM) began at the third week postoperatively and was limited at 0–90° for the next 2 weeks, allowing active flexion/extension; full ROM was allowed from the sixth week postoperatively.

Anteroposterior and lateral radiographs of the knee were obtained every 2–4 weeks until achievement of bony union and again at final follow-up. Details of mechanism of injury, associated injuries, timing of surgery, time to fracture healing, postoperative complications, final knee ROM, and implant removal were recorded. Time requested to regain normal daily activity was also scheduled. In addition, knee function was evaluated according to Bostman and Lysholm/Tegner scores [1].

Results

Of 20 patients, 17 were available for clinical and radiological follow-up (12 men and 5 women). The main mechanisms of injury were fall from height on hard ground in ten cases (59 %) and road traffic accidents in the remaining seven (41 %). Average patient age at injury was 46.6 years (range 20–75). There were nine transverse fractures, six comminuted, and two of the inferior pole. The average follow-up was 33 months (range 8–68). No patient had significant flexion contracture. Average ROM was 131.1° for flexion (range 120°–140°) and 0.5° for extension (range 0°–3°). No significant ROM differences were noted with the uninjured contralateral knee. Average Bostman scores at 3 months postoperatively and final follow-up were 25.2 ± 2 (range 20–30) and 28.3 ± 1.6 (range 26–30), respectively. Further, the mean Lysholm score at final follow-up was 91 points ±5.7 (range 83–100).

All fractures healed (time to union 9.2 ± 2 weeks) and no fixation failure was observed in the group studied. Slight losses of reduction (<4 mm) were noted in two patients at 4 weeks postoperatively [19, 22]. Because of noncompliance with the postoperative rehabilitation protocol, one patient presented knee stiffness at 2 months postoperatively that required gentle manipulation under anesthesia (Fig. 2). The same patient was the only one in the study to requrie elective FiberWire removal 24 months after surgery. This was performed secondary to a superficial infection, which did not affect the final clinical outcome (ROM 0–125°). Another patient underwent knee arthroscopy in another Institution because of anterior knee pain. No patients referred localized pain deriving from prominent suture knots (Fig. 3). Demographics, fracture type, and outcomes are presented in Table 1.

Fig. 2
figure 2

Comminuted patellar fracture. Preoperative X-ray (a); postoperative X-ray (b); final follow-up X-ray (c)

Fig. 3
figure 3

Transverse patellar fracture. Preoperative X-ray (a); postoperative X-ray (b); final follow-up X-ray (c)

Table 1 Demographic data, clinical outcomes, and postoperative complications of patients who underwent open reduction and internal fixation using a tension-band-wiring technique with a FiberWire suture

Discussion

The most important finding of our study was that FiberWire tension-band technique showed positive clinical outcomes with low implant-related complication rates. Just one of 17 patients required elective implant removal. No other patient complained of symptomatic hardware. Further, all fractures healed, and no fixation failure was noted.

Different methods have been described for treating patellar fractures. Traditionally, transverse fractures are best treated with a tension-band technique using two axial K-wires, with a figure-of-eight wire placed anteriorly [18]. This technique could be combined with the use of screws in case of more comminuted fracture types [5]. However, stainless steel wire is a difficult material to manipulate through tissues and may result in poor fracture fixation [6]. Further, high rates of metal-implant-related complications have been reported, such as K-wire migrationand skin irritation due to prominent wire knots, a very common complication, ranging from 0 % to 60 % [4, 17], possibly necessitating metal implant removal. In a case series of 27 patients, LeBrun et al. reported a hardware removal rate of approximately 52% at a mean of 6.5 years of follow-up [15]. Recently, Hoshino et al. performed a retrospective study of surgically treated patellar fractures. In this study, elective implant removal was performed in 37 % and 23 % of patients treated respectively with K-wires and cannulated screws [10].

On the basis of these results, some authors have advocated the use of nonmetallic implants, such as nonabsorbable suture and biodegradable cannulated screws [6, 7, 9, 11, 21, 25]. Many of these studies reported good outcomes for patellar fractures, with very low rate of postoperative complications. However, the concept of replacing stainless steel wire with sutures remains controversial in the literature, mainly due to uncertainty regarding rigid fracture fixation. For this reason, some authors evaluated the biomechanical properties of sutures used for patellar fractures. Chatakondu et al. found a significantly lower tensile strength for Ti-Cron sutures compared with stainless steel wire (14.80 vs 34.91 kg) [6]. McGreal et al. reported that braided polyester suture is an acceptable alternative to wire in tension-band fixation after testing cadaveric patellae for >20,000 cycles of knee flexion and extension [16]. Also, Patel et al., comparing different techniques including polyester sutures (no. 5 Ethibond), concluded that braided polyester sutures were comparable to stainless-steel wire for transverse patellar in terms of quality of fixation [19]. The FiberWire represents a particular suture characterized by a core of several small, individual strands of ultra-high-molecular-weight polyethylene covered with braided polyester suture material. It is used in a variety of orthopedic procedures, such as quadriceps/patellar tendon repairs, ACL reconstruction, rotator cuff repairs, and Achilles tendon repairs. FiberWire was evaluated biomechanically for tension-band fixation of a transverse patellar fracture: Using a three-point-bend model, Wright et al. observed that a double-strand FiberWire presented a significantly higher failure load than stainless steel wire. Furthermore, it was observed that, unlike stainless steel, FiberWire maintained its initial stiffness until failure [24]. This was confirmed by our study: Using a no. 5 FiberWire tension band, we found no significant fracture displacement following knee mobilization. Only two patients presented slight losses of reduction (<4 mm). This could be justified by progressive adhesion and adjustment of the suture through the peripatellar tissue that could be present during simple load, such as quadriceps muscle contraction. Further, high losses of reduction (>4 mm) and synthesis failure, which are potentially linked to FiberWire failure or breakage, were not noted in our case series. In addition, all treated fractures healed at 3 months postoperatively. The only patient requiring FiberWire removal was a 22-year-old man who was noncompliant with the postoperative rehabilitation protocol. In fact, this patient required gentle manipulation under anesthesia because of knee stiffness at 2 months postoperatively. Further, the same patient underwent elective FiberWire removal secondary to a superficial infection, probably due to implant-related soft-tissue irritation.

On the basis of the results of this study, we found that FiberWire tension bands can be used in place of metal-wire tension bands to treat patellar fracture, reducing the rate of symptomatic hardware.

Our study has several weaknesses. Major limitations are its retrospective nature and small size. However, our group was similar in size to other studies reporting results of patellar fractures. Further, we believe that high-quality prospective, randomized studies are required to define the effectiveness of nonabsorbable suture for internal fixation of patellar fracture.

In conclusion, the study demonstrates that a modified tension band using FiberWire sutures showed satisfactory clinical results with low incidence of complications and reoperations. Thus, FiberWire tension bands could be considered an alternative solution for treating patellar fracture, thus reducing the rate of symptomatic hardware.

References

  1. Böstman O, Kiviluoto O, Nirhamo J (1981) Comminuted displaced fractures of the patella. Injury 13:196–202

    Article  PubMed  Google Scholar 

  2. Bostrom A (1972) Fracture of the patella. A study of 422 patellar fractures. Acta Orthop Scand Suppl 143:1–80

    Article  CAS  PubMed  Google Scholar 

  3. Carpenter JE, Kasman R, Matthews LS (1993) Fractures of the patella. J Bone Joint Sur Am 75:1550–1561

    Google Scholar 

  4. Carpenter JE, Kasman R, Matthews LS (1994) Fractures of the patella. Instr Course Lect 43:97–108

    CAS  PubMed  Google Scholar 

  5. Carpenter JE, Kasman RA, Patel N, Lee ML, Goldstein SA (1997) Biomechanical evaluation of current patella fracture fixation techniques. J Orthop Trauma 11:351–356

    Article  CAS  PubMed  Google Scholar 

  6. Chatakondu SC, Abhaykumar S, Elliott DS (1998) The use of nonabsorbable suture in the fixation of patellar fractures: a preliminary report. Injury 29:23–27

    Article  CAS  PubMed  Google Scholar 

  7. Chen CH, Huang HY, Wu T, Lin J (2013) Transosseous suturing of patellar fractures with braided polyester—a prospective cohort with a matched historical control study. Injury 44:1309–1313

    Article  CAS  PubMed  Google Scholar 

  8. Curtis MJ. Internal fixation for fractures of the patella (1990) A comparison of two methods. J Bone Joint Surg Br 72:280–282

    Google Scholar 

  9. Gosal HS, Singh P, Field RE (2001) Clinical experience of patellar fracture fixation using metal wire or nonabsorbable polyester—a study of 37 cases. Injury 32:129–135

    Article  CAS  PubMed  Google Scholar 

  10. Hoshino CM, Tran W, Tiberi JV et al (2013) Complications following tension-band fixation of patellar fractures with cannulated screws compared with kirschner wires. J Bone Joint Surg Am 95:653–659

    Article  PubMed  Google Scholar 

  11. Hughes SC, Stott PM, Hearnden AJ, Ripley LG (2007) A new and effective tension-band braided polyester suture technique for transverse patellar fracture fixation. Injury 38:212–222

    Article  PubMed  Google Scholar 

  12. Hung LK, Chan KM, Chow YN, Leung PC (1985) Fractured patella: operative treatment using the tension band principle. Injury 16:343–347

    Article  CAS  PubMed  Google Scholar 

  13. John J, Wagner WW, Kuiper JH (2007) Tension-band wiring of transverse fractures of patella. The effect of site of wire twists and orientation of stainless steel wire loop: a biomechanical investigation. Int Orthop 31:703–707

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  14. Kumar G, Mereddy PK, Hakkalamani S, Donnachie NJ (2010) Implant removal following surgical stabilization of patella fracture. Orthopedics 12:33

    Google Scholar 

  15. LeBrun CT, Langford JR, Sagi HC (2012) Functional outcomes after operatively treated patella fractures. J Orthop Trauma 26:422–426

    Article  PubMed  Google Scholar 

  16. McGreal G, Reidy D, Joy A, Mahalingam K, Cashman WF (1999) The biomechanical evaluation of polyester as a tension band for the internal fixation of patellar fractures. J Med Eng Technol 23:53–56

    Article  CAS  PubMed  Google Scholar 

  17. Melvin JS, Mehta S (2011) Patellar fractures in adults. J Am Acad Orthop Surg 19:198–207

    Article  PubMed  Google Scholar 

  18. Muller ME, Allgower M, Schneider R, Willenegger H (1979) Manual of internal fixation. Techniques recommended by the AO-ASIF group. Springer, Berlin, pp 248–253

    Book  Google Scholar 

  19. Patel VR, Parks BG, Wang Y, Ebert FR, Jinnah RH (2000) Fixation of patella fractures with braided polyester suture: a biomechanical study. Injury 31:1–6

    Article  CAS  PubMed  Google Scholar 

  20. Petrie J, Sassoon A, Langford J (2013) Complications of patellar fracture repair: treatment and results. J Knee Surg 26:309–312

    Article  PubMed  Google Scholar 

  21. Qi L, Chang C, Xin T et al (2011) Double fixation of displaced patella fractures using bioabsorbable cannulated lag screws and braided polyester suture tension bands. Injury 42:1116–1120

    Article  PubMed  Google Scholar 

  22. Smith ST, Cramer KE, Karges DE, Watson JT, Moed BR (1997) Early complications in the operative treatment of patella fractures. J Orthop Trauma 11:183–187

    Article  CAS  PubMed  Google Scholar 

  23. Weber MJ, Janecki CJ, McLeod P, Nelson CL, Thompson JA (1980) Efficacy of various forms of fixation of transverse fractures of the patella. J Bone Joint Surg Am 62:215–220

    CAS  PubMed  Google Scholar 

  24. Wright PB, Kosmopoulos V, Cote RE, Tayag TJ, Nana AD (2009) FiberWire is superior in strength to stainless steel wire for tension band fixation of transverse patellar fractures. Injury 40:1200–1203

    Article  CAS  PubMed  Google Scholar 

  25. Yotsumoto T, Nishikawa U, Ryoke K, Nozaki K, Uchio Y (2009) Tension band fixation for treatment of patellar fracture: novel technique using a braided polyblend sutures and ring pins. Injury 40:713–717

    Article  PubMed  Google Scholar 

Download references

Ethical standards

All patients gave informed consent before inclusion into the study. The study was authorized by the local ethical committee and was performed in accordance with the ethical standards of the 1964 Declaration of Helsinki as revised in 2000.

Conflict of interest

The authors declare that they have no conflict of interest related to the publication of this manuscript.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Lawrence Camarda.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Camarda, L., La Gattuta, A., Butera, M. et al. FiberWire tension band for patellar fractures. J Orthopaed Traumatol 17, 75–80 (2016). https://doi.org/10.1007/s10195-015-0359-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10195-015-0359-6

Keywords