Table 3 Peak tension in the ulnar side on different configurations

From: Study of a constrained finite element elbow prosthesis: the influence of the implant placement

 

Resulting force (N)

Peak tension in the ulnar cement (MPa)

A: − 3°/0°

261.53

22.387

B:  + 3°/0°

254.55

29.395

C: 0°/− 3°

254.27

33.381

D: 0°/ + 3°

254.47

34.152