Skip to main content

Official Journal of the Italian Society of Orthopaedics and Traumatology

Table 1 Measured micromotion (μm) at baseline and at 10,000 cycles for each test specimen

From: Glenoid baseplate fixation using hybrid configurations of locked and unlocked peripheral screws

Test block

Hybrid screw fixation group

1-locked

2-locked

3-locked

4-locked

Baseline

10,000 cycles

Baseline

10,000 cycles

Baseline

10,000 cycles

Baseline

10,000 cycles

1

56.2

123.7

58.8

143.1

33.6

39.5

55.8

68.1

2

36.9

40.1

37.0

54.2

51.1

71.3

49.0

78.6

3

68.8

89.1

40.3

43.1

42.4

49.9

42.2

53.7

4

70.2

124.7

51.5

84.5

61.4

121.6

48.5

91.0

5

53.7

177.6

45.7

74.4

62.4

96.8

56.9

73.7

6

52.1

61.0

47.7

89.7

37.8

49.1

38.5

45.2

7

49.9

68.7

47.4

47.6

43.5

78.6

38.4

66.5