Table 4 Summary of studies reporting gross trunnion failure

From: Trunnionosis in total hip arthroplasty: a review

References

Number of hip implants

Implant description

Results/findings

Banerjee et al. [2]

5

Five different stem designs

Trunnions exhibited either a fracture or gross loss of material volume upon revision

Hohman et al. [34]

1

First-generation proximally coated titanium cementless stem

Damaged liner insert, an intact femoral head, diffuse metallic debris, and excessive wear of the trunnion

Pansard et al. [35]

1

12-mm/14-mm trunnion

Abnormal mobility between the femoral head and the trunnion, with severe damage at the taper. Macroscopic pitting and wear

Botti et al. [36]

1

Modular locking femoral prosthesis

Trunnion fracture. Pitting and crevice corrosion on fracture surface. Black metallic debris on proximal femoral stem